SURGICAL SPECIALTIES CORPORATION QUILL; 2-0 UNI MONODERM 20CM Back to Search Results
Model Number VLM-2005
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported there was a needle and thread separation during a surgical procedure.In one case of a tapp procedure, the needle separated and got lost in the body cavity, which resulted in a prolonged duration of the procedure.This is one of two devices that were reported.It is not certain with which device the event occurred.Reference complaint (b)(4).
 
Additional Manufacturer Narrative
A review of the device history records for the reported finished good lot and raw material components was performed and no quality issues were identified.Sterile samples from the reported lot were not returned for visual review and testing.The actual devices or magnified photos were not returned for review.If samples or photos become available at a later time, they will be reviewed and tested and the results will be included in the file.A follow-up report will be submitted at that time.A potential root cause for a needle detaching when slightly tugged during a procedure could be that the suture was not fully inserted within the end of the needle during the manufacturing process.It¿s also possible the device(s) are being grasped on or near the swaged end, damaging the suture, allowing it to break free from the needle component.The ¿precautions¿ section in the ifu for the device states, ¿care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point¿.Without reviewing the finished good lot number to review manufacturing records, receiving the actual reported device(s), receiving magnified photos of the actual devices, or receiving detailed information regarding the method utilized to remove the device from the packaging, preoperative preparation of the device, tools utilized to grasp the devices, details of the procedure performed, or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUILL
Common Device Name
2-0 UNI MONODERM 20CM
Manufacturer (Section D)
SURGICAL SPECIALTIES CORPORATION
corredor tijuana rosarito 20
#24702 b
tijuana baja california 22235
MX  22235
Manufacturer (Section G)
SURGICAL SPECIALTIES CORPORATION
247 station drive
westwood 02090
Manufacturer Contact
virnalisa cobacha
corredor tijuana-rosarito 2000 #24702b
ejido francisco villa
tijuana 22235
MX   22235
MDR Report Key19470857
Report Number3010692967-2024-00037
Device Sequence Number3554476
Product Code GAM
UDI-Device Identifier10848782020482
UDI-Public(01)10848782020482(17)281028(10)DA25JHC
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K141778
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/05/2024
Report Date (Section F) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVLM-2005
Device Lot NumberDA25JHC
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer05/20/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2024
Initial Report FDA Received Date06/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-