|
Combination product: yes.Upon receipt, the lead under complaint was found cut 13.5 cm distal to the is-1 connector pin (into 2 segments).All fragments were received for analysis.An extraction aid was found on the distal fragment.The insulation was found rubbed through approximately 9 cm proximal to the lead tip, which is assumed to be the root cause of the reported oversensing.Based on the characteristics of this damage, it is reasonable to assume that the lead had been subject to severe mechanical stress in the implanted state.Significant lead motion in the area of the tricuspid valve and interaction with atypical tissues should be taken into consideration.Furthermore, the rv shock coil, inner conductor coil in the distal end region and fixation helix were found stretched.These deformations probably occurred during the extraction procedure due to tensile stress.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|