| Catalog Number |
67PFSS30 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that multiple tracheostomy tubes once processed by csr are being returned to unit with "fuzzy" substance on obturator.There was patient involvement and no patient harm.No medical intervention required.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the sample returned consists in one (1) unit for p/n: 67pfss30 and l/n: 4400641, returned sample was received in used condition, decontaminated and in a plastic bag.The sample was visually inspected, at 12¿ to 16¿ under normal conditions of illumination.No damage or foreign particles were found in the returned sample; the obturator is clean without fuzzy.The sample is in good condition.The customer's failure mode was not observed.Based on the analysis performed on the sample provided and the reported by the customer, no root cause could be determined since the complaint was not confirmed.The failure mode will continue to be monitored and if a trend is identified an investigation will be opened.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|