The reported event could not be confirmed for the poly component due to the lack of additional clinical information.Upon further investigation of the ct scans by healthcare professionals the following was observed: there is information given, that a revision is planned for pain.Otherwise, we do not have any further clinical information regarding this case.The ct scan shows radiolucency and a very large cyst at the site of the tibial component.Loosening can be confirmed, there is no clear sign of migration.The talar component shows cysts as well.There is no clear sign of loosening or migration.Overall, nothing can be said regarding the underlying cause of the loosening and the cysts, although poor bone substance seems a likely cause.¿failure of total ankle replacement (tar) over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient (radiological and clinical) information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information (e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician) is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and/or the device in relation to cause of the failure.¿ although the issue is most likely related to poor bone substance, the root cause could not be conclusively determined.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|