The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the control body fluid damage, the down pulley wire fluid damage, the lg connector fluid damage, the light guide cable coating damage, the rubber trim collar broken, the control body corroded, the lg connector corroded, the light guide cable leak, the insertion flexible tube cut, the down pulley wire worn out, the lcb (light carrying bundle) defective, and the ccd module with drive pcb distorted image; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|