| Model Number |
MA60AC |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A non-healthcare professional reported with a description of distorted intraocular lens (iol).Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.No similar complaint and no non-conformance reported for the product lot/batch/serial under investigation.Associated products were not provided.It is unknown if qualified products were used.The company lens model is only qualified for use in the company ii (a) and (b) cartridges.The product investigation could not identify a root cause for the reported complaint.The product was not returned.It is unknown if qualified associated products were used.The company lens model is only qualified for use in the company ii (a) and (b) cartridges.The instruction for use (ifu) instructs: company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|