|
Fill volume: unknown.Flow rate: unknown.Procedure: unknown.Cathplace: unknown.Infusion start time: (b)(6) 2024.Infusion stop time: unknown.It was reported, patient reports a fast flow with homepump c series.Pump was covered with a blanket during treatment but the pump was placed on top of the cover, flow restrictor was not taped to the patient's body; approximately 5 hours left of treatment when pump was empty.Patient repots no injury occurred.Treatment was done on (b)(6) 2024.
|
|
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided; therefore, the udi-pi is unavailable.A root cause could not be determined.All information reasonably known as of 06 jun 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|