MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 3 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 3 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-35-3013
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Laxity (4526); Implant Pain (4561)
Date of Event 08/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitant devices: 2094326 02-012-41-3030 - logic tibia traptray cem sz 3f/3t 2167718 02-010-01-0330 - logic femoral ps cem right sz 3 2188938 200-02-32 - three peg patella 32mm the product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was the cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 141 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, anterior knee pain, instability in the knee with varus-valgus stress and hyperextension, worn and undersized patellar component, everted patella, and mild wear on the tibial insert with some fissuring and cracking noted.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-01816.This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: h6.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 3 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19479671
Report Number1038671-2024-01815
Device Sequence Number7072755
Product Code JWH
UDI-Device Identifier10885862001825
UDI-Public10885862001825
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/10/2018
Device Catalogue Number02-012-35-3013
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/28/2025
Initial Report FDA Received Date06/06/2024
Supplement Report FDA Received Date03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/11/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-