• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROHO, INC. NUMOTION CUSHION (REGISTERED AS ROHO QUADTRO SELECT HIGH PROFILE CUSHION); WHEELCHAIR CUSHION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROHO, INC. NUMOTION CUSHION (REGISTERED AS ROHO QUADTRO SELECT HIGH PROFILE CUSHION); WHEELCHAIR CUSHION Back to Search Results
Model Number NMSS1110C
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Cellulitis (1768); Pressure Sore/Ulcer (2326)
Date of Event 04/30/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer claims that she developed pressure sore and cellulitis due cushion going flat.Customer claims she sought medical attention at urgent care and wound clinic, but roho, inc.Has not seen medical records to confirm this.The manufacturing records were reviewed and indicated the cushion passed all inspections and verifications.The user has warning in the manual to check skin frequently that reads as follows: "check skin frequently, at least once a day.Redness, bruising, or darker areas (when compared to normal skin) may indicate superficial or deep tissue injury and should be addressed.If there is any discoloration to skin/soft tissue, stop use immediately.If the discoloration does not disappear within 30 minutes after disuse, immediately consult a healthcare professional.Check inflation frequently, at least once a day.Do not use a product that is underinflated or overinflated, because 1) the product benefits will be reduced or eliminated, resulting in an increased risk to skin and other soft tissue." the cushion was returned and an evaluation report is attached.No problem found during evaluation of the returned cushion.Issue determined to be caused by usage error.If roho, inc.Becomes aware of additional, relevant information, a follow-up report will be submitted.
 
Event or Problem Description
Customer claims she developed pressure sore and cellulitis due to cushion going flat.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUMOTION CUSHION (REGISTERED AS ROHO QUADTRO SELECT HIGH PROFILE CUSHION)
Common Device Name
WHEELCHAIR CUSHION
Manufacturer (Section D)
ROHO, INC.
1501 s 74th st
belleville IL 62223
Manufacturer (Section G)
ROHO, INC.
1501 s 74th st
belleville IL 62223
Manufacturer Contact
cynthia jackson
1501 s 74th st
belleville, IL 62223
6182223542
MDR Report Key19480689
Report Number3008630266-2024-00003
Device Sequence Number4072119
Product Code KIC
UDI-Device Identifier00613732084806
UDI-Public(01)00613732084806
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 06/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberNMSS1110C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2024
Initial Report FDA Received Date06/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-