| Model Number |
FIBERTAK DISPOSABLES STRAIGHT KIT |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the device during use.As no further investigation was able to be performed no change in harm was identified.
|
| |
|
Event or Problem Description
|
|
It was reported on 09/19/2022 by a sales representative via email that an ar-3638ds fibertak kits drill bit shaft seized up in the drill guide and broke.This was discovered during a case with no patient harm.
|
| |
|