Model Number FIBERTAK DISPOSABLES STRAIGHT KIT
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the device during use.As no further investigation was able to be performed no change in harm was identified.
 
Event or Problem Description
It was reported on 09/19/2022 by a sales representative via email that an ar-3638ds fibertak kits drill bit shaft seized up in the drill guide and broke.This was discovered during a case with no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAK DISPOSABLES STRAIGHT KIT
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19481370
Report Number1220246-2024-04973
Device Sequence Number18154653
Product Code LXH
UDI-Device Identifier00888867287440
UDI-Public00888867287440
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFIBERTAK DISPOSABLES STRAIGHT KIT
Device Catalogue NumberAR-3638DS
Device Lot Number14975328
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Initial Report FDA Received Date06/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-