Catalog Number 02-012-44-2513
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Discomfort (2330); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Joint Laxity (4526); Implant Pain (4561); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/03/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal documentation that approximately 67 months after a left total knee replacement procedure, the patient has experienced prosthesis wear, pain and discomfort, limited activities of daily living, impacted quality of life; debilitating injuries and damages, including but not limited to, pain and discomfort; swelling; gait impairment; poor balance; component part loosening; soft tissue damage; bone loss; bone damage; and other injuries presently undiagnosed, which all require ongoing medical care; future damages, including but not limited to cost of medical care; rehabilitation; home health care; loss of earning capacity; mental and emotional distress; and pain and suffering.No further issues or complications were reported.No additional information is available.Cases, (b)(4) is associated with this case.
 
Additional Manufacturer Narrative
D10 concomitant devices: 4862425 02-012-66-1000 - metaphyseal tibial cone, ml29mm 5130714 02-012-64-1212 - tru fluted stm ext 12mm x120mm blast 5160244 02-012-45-2515 - lgc tibial fit tray cem sz 2.5f / 1.5t 5394431 204-70-00 - tibial stem ext.Screw.The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the reported event cannot be confirmed from the information provided but may be due to prosthesis wear and/or inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA IMPLANT PSC INSERT, SZ 2.5, 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key19483263
Report Number1038671-2024-01804
Device Sequence Number8110213
Product Code JWH
UDI-Device Identifier10885862173539
UDI-Public10885862173539
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110547
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/29/2021
Device Catalogue Number02-012-44-2513
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2024
Supplement Date Received by Manufacturer03/14/2025
Initial Report FDA Received Date06/07/2024
Supplement Report FDA Received Date03/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/19/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexFemale
-
-