| Model Number |
M00550350 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was attempted to be used in the pulmonary tracts during a bronchoscopy procedure performed on (b)(6) 2024.During the procedure, the balloon had a hole in the middle of a surgical procedure.The procedure was completed with another cre pulmonary dilatation balloon.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a041001 captures the reportable event of a balloon pinhole.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a041001 captures the reportable event of a balloon pinhole.Block h11: investigation results: the returned cre pulmonary dilatation balloon was analyzed, and a visual inspection found no damages to the device.Functional and microscopic inspection found that the balloon had a pinhole located approximately at 15 mm from the tip of the device.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon pinhole was confirmed.The pinhole found is likely to have occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during or previous to the procedure could have caused the problem found on the distal section of the balloon.Therefore, the most probable root cause is an adverse event related to procedure.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was attempted to be used in the pulmonary tracts during a bronchoscopy procedure performed on (b)(6) 2024.During the procedure, the balloon had a hole in the middle of a surgical procedure.The procedure was completed with another cre pulmonary dilatation balloon.There were no patient complications reported as a result of this event.
|
| |
|