| Model Number |
FIBERTAK DISPOSABLES STRAIGHT KIT |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2023 it was reported by a sales representative via sems that (qty.2) ar-3638ds disposable drill kits would have stick while pulling out around halfway out.Both kits were discarded after the case.This was discovered during a procedure with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to a manufacturing issue; however, due to the device not being returned, we are unable to confirm the reported event.
|
| |
|