This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The customer's allegation was confirmed.Three attempts were performed to obtain additional information, but no response was received from the customer.Based on the results of the investigation, it is likely that the following led to the malfunction: physical stress applied to the insertion section during user handling, causing abnormality in the image sensor unit.A definitive root cause cannot be identified.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: "important information ¿ please read before use precaution: caution: turn the video system center on only when the endoscope connector is connected to the video system center.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.¦precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system ¦inspection of the endoscopic image confirm that the wli, nbi, and rdi endoscopic images are normal.5.1 troubleshooting if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿." olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|