Catalog Number 33654307
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/15/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.When additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the patient underwent a total ankle revision surgery to remove the implants and replace with a cement spacer due to infection.The patient didn't have the normal amount of post-op appointments due to the covid lockdown.
 
Event or Problem Description
It was reported that the patient underwent a total ankle revision surgery to remove the implants and replace with a cement spacer due to infection.The patient didn't have the normal amount of post-op appointments due to the covid lockdown.
 
Additional Manufacturer Narrative
The reported event could not be confirmed due to the lack of additional clinical information.Upon further investigation of the ct scans by healthcare professionals the following was observed: from the information provided, we know that the date of implant was (b)(6) 2020 and date of explant was (b)(6) 2024.The ct scan shows a girdlestone situation without any further information.¿failure of total ankle replacement (tar) over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient (radiological and clinical) information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information (e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician) is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and/or the device in relation to cause of the failure.¿ a microbiologist reviewed the sterilization procedures, environmental monitoring, bioburden data and the dhr and noted: the subject device was packaged according to established design and process specifications and was sterilized according to procedure.No deviations or non-conformances could be found.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY POLY SZ 3+ 7MM TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19484706
Report Number3010667733-2024-00353
Device Sequence Number12739015
Product Code HSN
UDI-Device Identifier00840420122742
UDI-Public00840420122742
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K140749
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/27/2025
Device Catalogue Number33654307
Device Lot Number1603097
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2024
Supplement Date Received by Manufacturer07/30/2024
Initial Report FDA Received Date06/07/2024
Supplement Report FDA Received Date08/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/29/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexUnknown
-
-