• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ARTHREX ECLIPSE HUMERAL HEAD, 47/18; TOT ANAT SHOULDR PROSTH, UNCEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ARTHREX ECLIPSE HUMERAL HEAD, 47/18; TOT ANAT SHOULDR PROSTH, UNCEM Back to Search Results
Model Number ARTHREX ECLIPSE HUMERAL HEAD, 47/18
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/12/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after an initial shoulder prosthesis surgery on (b)(6) 2020 the patient realized, in summer 2021, slight cracking noises in the left shoulder.During an examination in (b)(6) 2021 the orthopaedist of the patient identified changes on the x-rays.It was reported that a change of the plastic pan was identified.After two additional examinations by the orthopaedist it appears that the wear of the plastic pan als leads to a loosening of the pan.No further information were provided.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.  arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to a patient-specific event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHREX ECLIPSE HUMERAL HEAD, 47/18
Common Device Name
TOT ANAT SHOULDR PROSTH, UNCEM
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19484797
Report Number1220246-2024-05257
Device Sequence Number12743607
Product Code QHQ
UDI-Device Identifier00888867060043
UDI-Public00888867060043
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K183194
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model NumberARTHREX ECLIPSE HUMERAL HEAD, 47/18
Device Catalogue NumberAR-9347-18
Device Lot Number1910003
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Initial Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-