|
It was reported that the chief surgeon used an ultrasound knife to coagulate and cut the patient's tissue and blood vessels, however, the thunderbeat knife head was broken and unusable.The issue occurred during the therapeutic laparoscopic subtotal gastrectomy with gastrojejunostomy.The patient was under general anesthesia due to gastric cancer and pyloric obstruction.The procedure was completed with a replacement device, which delayed the procedure time by about 5 minutes.No fragment fell into the patient.There were no reports of patient harm.
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue the device was returned to olympus for inspection, and the customer reportable malfunction was confirmed.A definitive root cause was not identified, however based on the results of the investigation, the probable cause of the "broken probe" malfunction would likely be related to contact with other equipment, as follows: 1.During output activation in seal & cut mode, the probe came into contact with hard tissue, metal or surgical instruments.This caused scratches on the probe.2.A force to activate the output in seal & cut mode, or a force to grasp the body tissue was applied to the probe.Therefore, cracks were branching from the scratches on the probe.3.A force was applied to the probe causing it to break.The event can be prevented by following the instructions for use which state: ¿ the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, other instruments, or forceps, and others.Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/split/protruding/partial separating of the tissue pad.In turn, the probe tip may break before displaying an error window or generating an alarm tone.¿ do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.Olympus will continue to monitor field performance for this device.
|