| Model Number |
KREULOCK SCREW, TI, 3.5X16 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 11/9/2022, it was reported by a sales representative via email that an ar-8935cl-16 kreulock screw went right through an ar-9943t-07locking third tubular plate during insertion.This was discovered during an ankle fracture procedure on (b)(6) 2022.Surgeon backed the screw out a few turns and left it implanted in the patient.It is now acting a non-locking screw.
|
| |
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023. arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|