The manufacturer was previously received information regarding a simplygo mini device, returned to a third party service center for routine maintenance.During the routine maintenance device malfunction were repaired.The maintenance report indicated black residue in tubing, blackened air side.There is no allegation of harm or serious injury.Medical intervention was not reported.This notification was reviewed for pca-technical fault, comp.-black residue in tubing, sieves-compacted, air side-black end, o2 side-cracked, replaced faulty parts.The allegation has been reviewed by a service center and determined that the allegation of pca-technical fault, comp.-black residue in tubing, sieves-compacted, air side-black end, o2 side-cracked, replaced faulty parts does not qualify as a serious injury.Analysis of past events has not indicated an adverse event has occurred due to the reported allegation.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.This event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury and/or reportable product malfunction.
|