• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER SL 1 CURVE, 63 CM LENGTH, 8F; Introducer, catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER SL 1 CURVE, 63 CM LENGTH, 8F; Introducer, catheter Back to Search Results
Model Number 407439
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One 8.5f swartz braided introducer sheath was received for evaluation.The sheath had been perforated proximal to the distal tip.The sheath was flushed with fluid and the dilator was inserted into the sheath; no resistance or anomalies were noted.A needle/stylet assembly from current inventory was inserted and advanced through the dilator/sheath assembly; however, the needle could not advance past the location of the perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the needle insertion difficulty and perforation remains unknown.
 
Event or Problem Description
This report is to advise of a puncture found in the sheath during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER SL 1 CURVE, 63 CM LENGTH, 8F
Common Device Name
Introducer, catheter
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key19492173
Report Number3005334138-2024-00224
Device Sequence Number17901602
Product Code DYB
UDI-Device Identifier05414734205634
UDI-Public(01)05414734205634(10)10118430(17)261231
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K052644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number407439
Device Lot Number10118430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2024
Initial Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-