| Model Number |
407439 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One 8.5f swartz braided introducer sheath was received for evaluation.The sheath had been perforated proximal to the distal tip.The sheath was flushed with fluid and the dilator was inserted into the sheath; no resistance or anomalies were noted.A needle/stylet assembly from current inventory was inserted and advanced through the dilator/sheath assembly; however, the needle could not advance past the location of the perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the needle insertion difficulty and perforation remains unknown.
|
| |
|
Event or Problem Description
|
|
This report is to advise of a puncture found in the sheath during analysis.
|
| |
|