| Model Number |
7501349 |
| Medical Device Problem Codes |
Circuit Failure (1089); No Flow (2991); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the flushing pump exhibited a faulty pump head assembly and the power switch's led flickered.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation found the power switch's led flickered and the device failed to drive water through tubing due to faulty pump head assembly.Based on the results of the investigation, it is likely that the following led to the malfunction: the most probable cause of the complaint is cause traced to component failure, which is expected or random component failure without any design or manufacturing issue.A device history record review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|