• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION RING 620RG25 DURAN ANCORE ANNULOPLASTY C; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION RING 620RG25 DURAN ANCORE ANNULOPLASTY C; RING, ANNULOPLASTY Back to Search Results
Model Number 620RG25
Medical Device Problem Codes Insufficient Device Problem Information (3190); Device Stenosis (4066); Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Mitral Valve Stenosis (1965); Mitral Valve Insufficiency/ Regurgitation (4451); Insufficient Information (4580)
Date of Event 04/22/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Medtronic received information that 16 years and 3 months post implant of this 25mm mitral annuloplasty ring, it was explanted and replaced with a 29mm bioprosthetic valve.The reason for the replacement was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received additional information which stated that the reason for the replacement was recurrent moderate mitral regurgitation (mr) with moderate stenosis.It was also noted that a tricuspid valve repair was also carried out during this procedure.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RING 620RG25 DURAN ANCORE ANNULOPLASTY C
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19492548
Report Number2025587-2024-03321
Device Sequence Number7037203
Product Code KRH
UDI-Device Identifier00721902859762
UDI-Public00721902859762
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K032810
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/14/2012
Device Model Number620RG25
Device Catalogue Number620RG25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2024
Supplement Date Received by Manufacturer06/11/2024
Initial Report FDA Received Date06/07/2024
Supplement Report FDA Received Date06/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient Weight54 KG
-
-