| Catalog Number |
8065752885 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).Reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
An ophthalmologist reported that occlusion occurred continuously and an ophthalmic irrigation/aspiration handpiece did not aspirate in the visco phase of cataract surgery (without intra ocular lens implant).The surgery was completed after changing the cassette on the same day.There was no patient contact with the product.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections h.6 and h.11.A sample was not received at the manufacturing site for evaluation for the report of a dark area in the center; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review, complaint history review, and non-conformance review could not be conducted.A sample was not received at the manufacturing site and no lot information is available; therefore, the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All handpieces are 100% visually inspected prior to being released from the manufacturing facility.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|