| Model Number |
6996SQ58 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 4968-25 lead, (b)(6) 2017 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) subcutaneous defibrillation lead exhibited oversensing, undersensing, noise, and a l ead integrity alert (lia) was triggered by sensing integrity counter (sic) and high rate non-sustained episodes.It was further reported that the lead was plugged into the pacing port.A new rv pacing lead was implanted, and the rv subcutaneous defibrillation lead was moved to the defibrillation port and remains in use for high voltage therapy.No patient complications have been reported as a result of this event.
|
| |
|