Model Number 9.0 MM FLIP CUTTER II
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 3/6/2023, it was reported by a distributor via email that an ar-1204af-90 flipcutter ii broke while trying to flip it back after reaming the femoral tunnel.The case was completed using an ar-1204af-100.This was discovered during an alc procedure to repair the complete anterior cruciate ligament rupture on (b)(6) 2023.Additional information received on 3/7/2023: while using the ar-1204af-90 flipcutter ii, the blade would not close to its original state.It did not break off or have any loose fragments inside the patient.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the attachment during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9.0 MM FLIP CUTTER II
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19495245
Report Number1220246-2024-05336
Device Sequence Number12914602
Product Code HWE
UDI-Device Identifier00888867004269
UDI-Public00888867004269
Combination Product (Y/N)N
Initial Reporter CountryPE
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9.0 MM FLIP CUTTER II
Device Catalogue NumberAR-1204AF-90
Device Lot Number2187123267
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2023
Initial Report FDA Received Date06/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-