The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the air/water socket clogged.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the air/water socket.In addition, our technician confirmed that the water nozzle clogged, and the insertion flexible tube leak; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0630(air/water & jet water channels)" and/or the risk analysis results, it was evaluated to submit mdr.
|