Investigation: visual inspection: during the investigation, no significant deformation or damage of the valve, but some visible deposits in the reservoir could be determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve does not operate within the accepted tolerance in either position.An accelerated outflow of gav 2.0 could be determined.Internal inspection : after dismantling of the valve, visible deposits were found in gav 2.0.Results: based on our investigation results, we can determine an accelerated outflow in the valve.The determined deposits can be named as the cause for the accelerated outflow.Organic material in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
|