| Model Number |
3810002 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Slipped (1584); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation: a terumo bct service technician checked out the machine at the customer site.The technician adjusted the button for iv pole release, tested and returned to service.One year of service history was reviewed for this device with no further issues related to the reported condition identified.The device serial number history report indicates no further related issues have been reported for this device.Correction: a field service engineer adjusted the iv pole release button.Investigation is in process, a follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No adverse event occurred and no medical intervention was required.Iv pole clamp was not installed.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
|
| |
|
Event or Problem Description
|
|
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No adverse event occurred and no medical intervention was required.Iv pole clamp was not installed.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: a terumo bct service technician checked out the machine at the customer site.The technician adjusted the button for iv pole release, tested and returned to service.One year of service history was reviewed for this device with no further issues related to the reported condition identified.The device serial number history report indicates no further related issues have been reported for this device.Correction: a field service engineer adjusted the iv pole release button.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be related to the need for an iv pole release button adjustment.
|
| |
|