MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BOOST BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BOOST BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SHPB370
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Pocket Erosion (2013)
Date of Event 05/14/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a breast augmentation revision with mentor memorygel boost breast implants 370cc and experienced capsular contracture baker grade iv on the right side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent removal on (b)(6) 2024.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: capsular contracture manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On jul 3, 2024, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the smo ro high pro boost gel370cc breast implant was found to have some bubbles within the gel.Bubbles in the gel can be originated with changes in pressure and temperature which can affect volume.The shell wall is permeable to air, therefore trapped air can form bubbles with changes in pressure or temperature.When left by themselves with no changes in temperature or pressure, bubbles usually dissipate over time, unless trapped by cured gel.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body's individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On jul 11, 2024, additional information was received indicating the patient also experienced device extrusion on the right side.The replacement device was identified as a mentor gel breast implant (serial number (b)(6).Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BOOST BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key19499576
Report Number1645337-2024-06960
Device Sequence Number7973164
Product Code FTR
UDI-Device Identifier00081317027058
UDI-Public00081317027058
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSHPB370
Device Lot Number9970303
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/23/2024
Supplement Date Received by Manufacturer07/03/2024
07/11/2024
Initial Report FDA Received Date06/10/2024
Supplement Report FDA Received Date07/09/2024
08/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-