MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 39MM; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 39MM; HIP INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 244000539
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary according to the information received, ¿it was reported that the crossback graters was reported for unknown reason.Did not happen in surgery".The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the device revealed that the cutting edged presents signs of wear consistent with repeated use and servicing the observed condition was identified as an end of life indicator.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection and functional testing were not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the quickset ace grater head would have contributed to a device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the crossback grater was reported for unknown reason.Did not happen in surgery.Additional event information: visual analysis of the device revealed that the cutting edged presents signs of wear consistent with repeated use and servicing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HEAD 39MM
Common Device Name
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19499977
Report Number1818910-2024-12649
Device Sequence Number6661729
Product Code HTO
UDI-Device Identifier10603295123866
UDI-Public10603295123866
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number244000539
Device Lot NumberSO2051459
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2024
Initial Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-