| Model Number |
TIGHTROPE, ABS, IMPLANT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The most likely cause of the failure is use error due to excessive force on the device.As no further investigation was able to be performed no change in harm was identified.The results of the investigation of this complaint did not change the harm identified during intake.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament graft link surgery during the insertion of the graft, the tigtrope suture of the device tore.Therefore, the graft had to be completely created from the start.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a different device.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|