Model Number VAL SCREW, TI, 2.4 X 12MM
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/11/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure, while the surgeon was using the mini cfs set, the surgeon has an issue with two of the 2.4x12mm val locking screws, ar-18724v-12.The surgeon drilled 3 holes in the v plate and when tried to tighten the val screws into the hole, they would not engage.The surgeon had to switch the two 2.4x12mm screws and replace them with the 2.4x11mm screw and a 2.4x12mm screw after they did not lock into multiple different holes of the plate.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.One unpackaged ar-18724v-12 locking screw (no batch number provided) was received for investigation.Visual inspection under microscope identified the presence of bone remnants lodged within the locking threads, inferior to the hex.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device when cleaning and disinfecting by user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAL SCREW, TI, 2.4 X 12MM
Common Device Name
BONE FIXATION PLATE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19500822
Report Number1220246-2024-05452
Device Sequence Number7870575
Product Code HRS
UDI-Device Identifier00888867314801
UDI-Public00888867314801
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K220937
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVAL SCREW, TI, 2.4 X 12MM
Device Catalogue NumberAR-18724V-12
Device Lot Number2685631
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2021
Initial Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/30/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-