Model Number R1054655TP
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, an "internal battery" code was found in the ventilator's downloaded error log.The device's power management board was replaced to address the issue.Additional findings not related to the malfunction/issue were the pollen filter, exhaust fan assembly, 43-micron filter were proactively replaced on the device.After replacing the parts on the system, the repair test and running test, battery and valve checks, and a cleaning of the device were performed.The device was found to be operational.
 
Additional Manufacturer Narrative
A ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, an "internal battery" code was found in the ventilator's downloaded error log.The device's power management board was replaced to address the issue.The power management board was returned to the philips investigation lab (pil) for evaluation.The system error log was reviewed, and the system error was confirmed.Pil tested the power management board and was not able to reproduce the same error that was found in the error log.The part was examined, and no visual defects were found.The power management board was tested and both batteries showed sign of charging.Pil confirmed that they could not determine the underlying issue for the failure that was found by service.Pil concluded the power management board was functioning as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key19502941
Report Number2518422-2024-35496
Device Sequence Number15430658
Product Code CBK
UDI-Device Identifier00606959033894
UDI-Public00606959033894
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberR1054655TP
Device Catalogue NumberR1054655TP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/08/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/10/2024
Supplement Report FDA Received Date07/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/16/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-