| Model Number |
D314DRM |
| Medical Device Problem Codes |
Failure to Charge (1085); Electrical /Electronic Property Problem (1198); Pacing Problem (1439); Delayed Charge Time (2586); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) possibly delivered inappropriate anti-tachycardia pacing (atp) for supra ventricular tachycardia (svt) when the device detected the episode as fast ventricular tachycardia (fvt).It was also noted that charge circuit timeout and charge circuit inactive alerts were triggered at end of service (eos).It was also noted that the charge time was longer than expected.The icd was explanted and replaced. no patient complications have been reported as a result of this event.
|
| |
|