| Model Number |
LF-DP |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that "coating of the insertion tube is peeled off" occurred due to stress of repeated use, external factors, or handling.The event can be detected/prevented by following the instructions for use which state: ¿chapter 3 preparation and inspection 3.2 preparation and inspection of the endoscope".Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the intubation fiberscope exhibited coating of light guide hose peeled off.There was no patient involvement.
|
| |
|