• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VITALIO; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VITALIO; PACEMAKER Back to Search Results
Model Number K273
Medical Device Problem Code Telemetry Discrepancy (1629)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/01/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this pacemaker exhibited invalid radiofrequency overuse condition.Further information was received that this device was explanted.No additional adverse patient effects were reported.This device has been received for analysis.
 
Event or Problem Description
It was reported that this pacemaker exhibited invalid radiofrequency overuse condition.Further information was received that this device was explanted.No additional adverse patient effects were reported.This device has been received for analysis.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.The baseline testing completed on the device found no failing conditions.All testing that was completed on the device passed or was not applicable, therefore no problem was detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALIO
Common Device Name
PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19506404
Report Number2124215-2024-35364
Device Sequence Number7452570
Product Code LWP
UDI-Device Identifier00802526516962
UDI-Public00802526516962
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S143
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/07/2016
Device Model NumberK273
Device Catalogue NumberK273
Device Lot Number101674
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2024
Supplement Date Received by Manufacturer10/10/2024
Initial Report FDA Received Date06/10/2024
Supplement Report FDA Received Date10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/13/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age79 YR
Patient SexFemale
-
-