SYNTHES GMBH SYNFRAME HALF RING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 387.337
Medical Device Problem Code Compatibility Problem (2960)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.E3: reporter is a j&j sales representative.H3, h4, h6: part # 387.337 synthes lot # 5l03142 supplier lot # 5l03142 release to warehouse date:, 28 sep 2020 supplier: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.According to the information received, "[the nurse was having issues assembling the synframe ring, as the screws are stripped.The nurse requested another set in order to proceed for the surgery.]" the photos were provided for review.The photo investigation didn't revealed that [387.337, synframe half ring] had unable to assemble.The provided evidence was not sufficient to confirm the reported event.Functionality issues cannot be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the [synframe half ring] would not contribute to the complained device issue.Based on the investigation findings, potential cause can't be determined and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that a nurse was having issues assembling a syn-frame ring.It was reported that the screws are stripped.The nurse requested another set in order to proceed for the surgery.This event occurred pre-op, and the patient was not impacted.This report is for one (1) synframe half ring.This is report 2 of 4 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that device was worn out from the head of the adjustable screw.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.Functional test was performed, the mating device was assembly however, the adjustable screw was unable to close due the worn out condition of the head.The overall complaint was confirmed as the observed condition of the synframe half ring would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Corrected data: g1: manufacturing site name and address h4: product code: 387.337; sub-component: 50131166; lot number: 5l03142; release to warehouse date: 14.Sep.2019; expiration date: na; supplier: na; manufacturing site: werk hagendorf.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNFRAME HALF RING
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19506657
Report Number8030965-2024-07284
Device Sequence Number19721340
Product Code LXH
UDI-Device Identifier10705034771069
UDI-Public(01)10705034771069
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number387.337
Device Lot Number5L03142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/14/2024
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date06/10/2024
Supplement Report FDA Received Date08/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SYNFRAME HALF RING.; UNK - SCREWS: SPINE-US.; UNK - SCREWS: SPINE-US.
Patient SexUnknown
-
-