• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. GALILEO® CAPTURING ROD; INTRAMED FIXATN ROD AND ACCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. GALILEO® CAPTURING ROD; INTRAMED FIXATN ROD AND ACCESS Back to Search Results
Model Number GALILEO® CAPTURING ROD
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint allegation confirmed based on the photo provided by the reporter.Visual inspection revealed the capturing rod nut was damaged.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.
 
Event or Problem Description
It was reported on 02/14/2023 by a sales representative via (b)(4) that an 5046-100 capturing rod, 5033-100 capturing rod, and 0468-100 t-handle broke during use.Case involvement, no patient effect.Nothing broke inside patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALILEO® CAPTURING ROD
Common Device Name
INTRAMED FIXATN ROD AND ACCESS
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19509377
Report Number1220246-2024-05533
Device Sequence Number7498439
Product Code HSB
UDI-Device Identifier00848665039177
UDI-Public00848665039177
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K202099
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGALILEO® CAPTURING ROD
Device Catalogue Number5033-100
Device Lot Number220097
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2023
Initial Report FDA Received Date06/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-