ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 13 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42522100813
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: medical product: unknown femoral component: catalog#ni, lot#ni; unknown tibial tray: catalog#ni, lot#ni g2: foreign: germany diligence is in process to determine whether the product is available for evaluation.The investigation is in progress.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent an initial total knee arthroplasty.Approximately one (1) year and eight (8) months post-implantation, it was reported that an incorrectly sized articular surface had been implanted.Due diligence is in progress for this event; to date no additional information or medical intervention has been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; g3; h2; h3; h6; h11.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Investigation results concluded that the reported event was due to failure of the operator to follow instructions.Per the device instructions for use, the system components are sized by matching the femoral and tibial component sizes, styles, and orientation on the product label to the information on the articular surface component label and/or product compatibility charts included in the package insert and surgical technique manual.Mismatching may result in poor surface contact and could produce pain, decrease wear resistance, produce instability of the implant, or otherwise reduce the service life of the implant.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.Reported event was unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that a patient underwent an initial total knee arthroplasty.Approximately one (1) year and eight (8) months post-implantation, it was reported that an incorrectly sized articular surface had been implanted.No abnormalities or complications have been reported and the product remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 13 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19509614
Report Number3007963827-2024-00219
Device Sequence Number16992770
Product Code MBH
UDI-Device Identifier00889024468443
UDI-Public(01)00889024468443(17)270426(10)65349175
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number42522100813
Device Lot Number65349175
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2024
Supplement Date Received by Manufacturer11/21/2024
Initial Report FDA Received Date06/11/2024
Supplement Report FDA Received Date11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Patient Age62 YR
Patient SexFemale
-
-