• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number M3-30A
Medical Device Problem Codes Material Integrity Problem (2978); Sharp Edges (4013)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.New information added to the following fields: h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the damage to the tip of the scope ( burnt and sharp) occurred due to improper handling of the device.The event can be detected/prevented by handling the device in accordance with the following instructions for use: study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects;" "keep the distal tip of any electrode, probe, laser fiber, or other ancillary device in the field of view at all times when active ( page 4).Olympus will continue to monitor field performance for this device.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was observed that during the device inspection, the hysteroscope distal tip was found to be burnt and sharp.There were no reports of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19509881
Report Number3011050570-2024-00224
Device Sequence Number3554680
Product Code HIH
UDI-Device Identifier00821925005679
UDI-Public00821925005679
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K890328
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberM3-30A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2024
Supplement Date Received by Manufacturer06/19/2024
Initial Report FDA Received Date06/11/2024
Supplement Report FDA Received Date06/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-