ARTHREX, INC. BIO-COMP SWVLK SP, 4.75X24.5MM; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number BIO-COMP SWVLK SP, 4.75X24.5MM
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a knee surgery the device blocked.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.**22-feb-2023 update (b)(6).Further information were provided that the device could not be turned anymore.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP SWVLK SP, 4.75X24.5MM
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19510763
Report Number1220246-2024-05554
Device Sequence Number3404695
Product Code MAI
UDI-Device Identifier00888867098039
UDI-Public00888867098039
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K151342
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBIO-COMP SWVLK SP, 4.75X24.5MM
Device Catalogue NumberAR-2324BCM-1
Device Lot Number15036916
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2023
Initial Report FDA Received Date06/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-