MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 8300AB19A
Medical Device Problem Codes Calcified (1077); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Dyspnea (1816); Fatigue (1849); Heart Failure/Congestive Heart Failure (4446); Swelling/ Edema (4577); Insufficient Information (4580)
Date of Event 05/15/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Event or Problem Description
It was reported that a 19mm 8300ab in the aortic position, underwent a valve-in-valve procedure after an implant duration of 7 years, 7 months due to unknown reason.The tavr was performed with a 20mm 9755rsl transcatheter valve.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Calcific degeneration is a common cause of bioprosthetic heart valve failures.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valve's hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards' tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprostheses from calcifying.Calcification is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The most likely cause is patient factors, including ckd.
 
Event or Problem Description
It was reported that a 19mm 8300ab in the aortic position, underwent a valve-in-valve procedure after an implant duration of 7 years, 7 months due to calification and stenosis.The patient presented with heart failure, dyspnea, le edema, and fatigue.The tavr was performed with a 20mm 9755rsl transcatheter valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irivne CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key19510848
Report Number2015691-2024-04473
Device Sequence Number19684286
Product Code LWR
UDI-Device Identifier00690103194470
UDI-Public(01)00690103194470(17)180823
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P150036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/23/2018
Device Model Number8300AB19A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2024
Supplement Date Received by Manufacturer06/18/2024
Initial Report FDA Received Date06/11/2024
Supplement Report FDA Received Date07/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/23/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexFemale
-
-