MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VISIONIST CRT-P; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VISIONIST CRT-P; IMPLANTABLE DEVICE Back to Search Results
Model Number U226
Medical Device Problem Codes Signal Artifact/Noise (1036); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system exhibited noise on both the right atrial (ra) and right ventricular (rv) channel.Additionally, both leads had low out-of-range pacing lead impedance measurements measuring less than 200 ohms.Both leads are programmed too unipolar.It was noted there was inappropriate mode switching on the ra lead.Lead was programmed temporarily to bipolar and noise was still present.Technical services discussed possible causes and provided programming options.At this time, the system remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST CRT-P
Common Device Name
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19510887
Report Number2124215-2024-35523
Device Sequence Number3455745
Product Code NKE
UDI-Device Identifier00802526559457
UDI-Public00802526559457
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P030005/S138
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/15/2022
Device Model NumberU226
Device Catalogue NumberU226
Device Lot Number701446
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2024
Initial Report FDA Received Date06/11/2024
Date Device Manufactured09/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
-
-