| Model Number |
NANONEEDLE SCOPE, 180MM |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an arthroscopy on the right knee different issues with the console have occurred.White sporadic flashes were visible on the camera image.The camera button of the nano needle did not work.No images were visible on the tablet.( black image and no export to usb or ipad was possible for a specific case) and the headup display could not be activated.There was no harm for patient, operator or third party reported.The surgery was finished successfully with the same device anyway.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to improper cleaning process used - thus causing moisture intrusion.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided ¿ which may include the returned device (if available), photographs, videos, event description, and any additional field input ¿ arthrex has determined the most likely cause.The reported failure is most likely due to a supplier or manufacturing issue, resulting in the loss of image display and control functionality during arthroscopy.
|
| |
|