• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. MENISCAL CINCH; POLYETHYLENE SYNTHETIC SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. MENISCAL CINCH; POLYETHYLENE SYNTHETIC SUTURE Back to Search Results
Model Number MENISCAL CINCH
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Since the complaint device was not returned, a physical evaluation of the device was not performed.However, the complaint allegation is not confirmed without the device being returned or any photos provided.The most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the device during use.
 
Event or Problem Description
It was reported that the device prematurely deployed.Both implants were removed from the patient and a second device was used to complete the procedure successfully.See source data attached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENISCAL CINCH
Common Device Name
POLYETHYLENE SYNTHETIC SUTURE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19514908
Report Number1220246-2024-05582
Device Sequence Number3506286
Product Code GAT
UDI-Device Identifier00888867033535
UDI-Public00888867033535
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K073149
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model NumberMENISCAL CINCH
Device Catalogue NumberAR-4500
Device Lot Number11622463
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2021
Initial Report FDA Received Date06/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-