• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA POSTERIOR REFERENCING SERIES 1 SIZER; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA POSTERIOR REFERENCING SERIES 1 SIZER; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was noted that when the new sizer was going to be placed in a kit, the etching and tick marks were mis-aligned.There was no patient involvement.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, h10, and h11.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned products identified component (b)(4) doesn¿t line up with the arrow etched on component (b)(4) when it locks on all three devices.Dimensional analysis of the arrow etching of all returned devices found it is non-conforming to print specifications.Device history record was reviewed and no discrepancies were found.The root cause of the reported issue is attributed to a manufacturing issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA POSTERIOR REFERENCING SERIES 1 SIZER
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19515246
Report Number0001822565-2024-01964
Device Sequence Number17961836
Product Code JWH
UDI-Device Identifier00889024632202
UDI-Public(01)00889024632202(11)231016(10)66164945
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K221479
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42509904020
Device Lot Number66164945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2024
Supplement Date Received by Manufacturer11/13/2024
Initial Report FDA Received Date06/11/2024
Supplement Report FDA Received Date11/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-