| Catalog Number |
3501254BC |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
On june 13, 2024, mentor evaluated a photo of the explanted device.Photo evaluation summary: upon visual evaluation of the image provided in the complaint no apparent damage or visual anomalies were observed.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made. as such, the investigation will be closed. if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date. this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report. if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation surgery with 125cc mentor memorygel breast implants.Post-operatively, the patient experienced bilateral baker grade iii capsular contracture, which was confirmed by a medical professional.As a result, the patient underwent bilateral removal surgery on (b)(6) 2024.No date of implantation was provided.This report is for the left implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on august 28, 2024: during visual evaluation no apparent damage or visual anomalies were observed on the smooth hpg, 125cc breast implant.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Patient initial is (b)(6).Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|