• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE UNITIZED SET SCR; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE UNITIZED SET SCR; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 199721001S
Medical Device Problem Codes Break (1069); Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: b3: unknown event date d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E1: initial reporter is j&j company representative h3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported a revision surgery occurred on (b)(6) 2024 due to screw displacement and migration.In the surgery, it was confirmed that sfx broke off as well as the reported screws.The deviated screw was replaced.The fenestrated screws were added for the extension on upper vertebrae.All of the set screw were replaced.Sfx was performed.The surgery was completed successfully.Patient outcome was stable the original surgery occurred on (b)(6) 2024 and was for a spinal fusion (l5/s) (l4-sai fusion using verse), for pseudarthrosis on (b)(6) 2024.Patient was stable.Original procedure was completed successfully with no surgical delay.It was noted post-operatively on an unknown date that the screws had migrated.This report involves one (1) 5.5 exp verse unitized set scr this is report 1 of 6 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device and pictures provided.Visual analysis of the returned sample and photos provided revealed that there were no damage or defect with the 5.5 exp verse unitized set scr, only was observed with signs of usage which could have been result of implantation and explantation.Broken condition was not observed on the physical device received and photos.Migration of the device was not confirmed as the photos provided do not show the alleged reported condition.A dimensional inspection for the 5.5 exp verse unitized set scr was not performed due to post manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the 5.5 exp verse unitized set scr was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): a manufacturing record evaluation was performed for the finished device product code # 199721001s, lot # l20340.It was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 29-nov-2023.Manufacturing site: jabil le locle.Expiry date: 31-oct-2028.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE UNITIZED SET SCR
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key19518314
Report Number1526439-2024-01653
Device Sequence Number8412138
Product Code NKB
UDI-Device Identifier10705034466149
UDI-Public(01)10705034466149
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K142185
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number199721001S
Device Lot NumberL20340
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/15/2024
Supplement Date Received by Manufacturer06/12/2024
Initial Report FDA Received Date06/12/2024
Supplement Report FDA Received Date07/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 EXP VERSE UNITIZED SET SCR.; UNK - LOCKING/SET SCREWS: EXPEDIUM SFX.; UNKNOWN SCREWS.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-