• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEGADYNE MEDICAL PRODUCTS, INC. E-Z CLEAN PNCL RKR; E-Z CLEAN PENCIL ROCKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEGADYNE MEDICAL PRODUCTS, INC. E-Z CLEAN PNCL RKR; E-Z CLEAN PENCIL ROCKER Back to Search Results
Catalog Number 0030
Medical Device Problem Code Arcing of Electrodes (2289)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 6/12/2024.An analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure with the appropriate parameters already modified (power in the generator), it is evident that it generates an unusual flame in the surgical field, the electrocoagulation configuration is reduced to avoid burns.In the patient.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E-Z CLEAN PNCL RKR
Common Device Name
E-Z CLEAN PENCIL ROCKER
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
NEW DEANTRONICS TAIWAN, LTD.
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key19518399
Report Number1721194-2024-00055
Device Sequence Number17650001
Product Code GEI
UDI-Device Identifier10614559100462
UDI-Public10614559100462
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K965054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0030
Device Lot Number2209008
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/28/2024
Initial Report FDA Received Date06/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Patient SexUnknown
-
-