• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 165820
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while inserting an 18 fr catheter in a patient with hyperalgesic urinary retention, the urine did return.The nurse inflated the balloon with 10ml of sterile water but when the nurse mobilized the catheter, it did not stay in place.When they tried the second time with 20ml of sterile water, the same incident occurred again.They tried several times injecting sterile water more slowly, changing the batch or inserting a 20fr probe, but they could not succeed.The patient also felt the balloon pop once.The user also carried out blank tests outside the patient and the balloon was also puncturing.After several attempts, they succeeded in inserting the urinary catheter.The customer used 165818ce and 165820ce with batches mygq1137 and myey1462 respectively.The packaging of other unknown batches was not preserved, and the probes are available from the pharmacy for expert appraisal.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley.Visual inspection of the sample noted the catheter balloon was inflated with 10 ml methylene blue solution (3 drops 1 percentage aq methylene blue per 100ml distilled water) and pinhole present on the balloon measuring (0.013") therefore, confirmed as pinhole would lead to deflation due to trauma.Root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be ¿imperfection in balloon due to process¿.However, there was insufficient information to confirm this potential root cause.A dhr review did not show any problems or conditions.The instructions for use were found adequate and state the following: "warning: do not use ointments or lubricants having a petrolatum base.They will damage silicone.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml of sterile water.Or for prefilled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, reseat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Single use only.Do not resterilize.This is a single use device.Do not resterilize any portion of this device.Reuse and or repackaging may create a risk of patient or user infection, compromise the structural integrity and or essential material and design characteristics of the device, which may lead to device failure, and or lead to injury, illness or death of the patient." this report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that while inserting an 18 fr catheter in a patient with hyperalgesic urinary retention, the urine did return.The nurse inflated the balloon with 10ml of sterile water but when the nurse mobilized the catheter, it did not stay in place.When they tried the second time with 20ml of sterile water, the same incident occurred again.They tried several times injecting sterile water more slowly, changing the batch or inserting a 20fr probe, but they could not succeed.The patient also felt the balloon pop once.The user also carried out blank tests outside the patient and the balloon was also puncturing.After several attempts, they succeeded in inserting the urinary catheter.The customer used 165820ce, and 165818ce with batches mygq1137 and myey1462.The packaging of other unknown batches were not preserved, and the probes are available from the pharmacy for expert appraisal.Per sample return on (b)(6) 2024, the received sample did not match with the reported material number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE FOLEY CATHETER
Common Device Name
SILICONE FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19518532
Report Number1018233-2024-03444
Device Sequence Number20917062
Product Code EZL
UDI-Device Identifier00801741029776
UDI-Public(01)00801741029776(11)190731(17)240603(10)NGDT0366
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/03/2024
Device Catalogue Number165820
Device Lot NumberNGDT0366
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2024
Supplement Date Received by Manufacturer09/19/2024
Initial Report FDA Received Date06/12/2024
Supplement Report FDA Received Date09/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-